Page 1 - User Quick

MAGPIX® User QuickGuideFor In Vitro Diagnostic Use

Page 2

CAUTION: Correct sample probe height is critical to successful sampleacquisition and calibration. Problems with the sample probe canlead to fluid leak

Page 3

1. Open the Maintenance page, then the Cmds & Routines tab.2. Select Revive After Storage (Luminex) from the drop-down list. The Revive AfterStora

Page 4

System InitializationSystem initialization prepares the system for data acquisition. On initial startup, the SystemInitialization should consist of:•

Page 5 - Table of Contents

1. Load the CAL/VER CD (provided with the kit) on the computer.2. On the Home page, click System Initialization.3. Click Import Kit at the bottom righ

Page 6

MAGPIX® User Quick Guide8For In Vitro Diagnostic Use

Page 7 - Start-up Indicators

Defining the System Initialization RoutineNOTE: Luminex recommends daily performance verification and weeklycalibration of the MAGPIX system. You can

Page 8 - Initial Startup

1. On the Home page, click System Initialization under Daily Activities. The Auto Mainttab opens.2. On the Auto Maint tab, the System Initialization o

Page 9

Creating a KitTo create a kit:1. Open the Protocols page, then open the Protocols tab. Select the protocol that youwant to use for the kit, then click

Page 10 - Revive After Storage Routine

1. Open the Protocols page, then open the Protocols tab. Click the Stds & Ctrls tab, thenclick Create New Std/Ctrl Lots.2. In the Select Protocol

Page 11 - For In Vitro Diagnostic Use

5. If you want a description for the batch, type it in the Enter Optional Description box.• Click the existing protocol that you want to use. If the p

Page 13

6. Click Generate. The report appears in the lower part of the Reports tab.7. Click Print to print the report, or Save to save the report as a PDF fil

Page 14 - User Quick Guide

Shutting Down the AnalyzerRun the daily shutdown routine to prevent clogs and crystallization of salt in the sampleprobe. Clogs and crystallization of

Page 15

MAGPIX® User Quick Guide16For In Vitro Diagnostic Use

Page 16

MAGPIX New System Initialization TroubleshootingThis section provides you with information to troubleshoot issues that may be encounteredduring the MA

Page 17 - Chapter 3: Running Assays

1. Adjusting the Sample Probe Height2. Revive After Storage Routine (Luminex)3. System Initialization All system startup routines are run from the MAG

Page 18

Problem Summary Resolution NotesIf Calibration continues to fail runthe following routine:Create routines in thexPONENT “Maintenance” Tab“Cmds & R

Page 19 - Run, Save, and Print Reports

Problem Summary Resolution Notes(1x) Sanitize with 0.1N NaOH(reservoir C)(2x) Clean with 0.1N NaOH(reservoir C)(2x) Rinse (reservoir D)(1x) Alcohol Fl

Page 20 - Data Interpretation Report

FIGURE 7.MAGPIX Calibration and Performance Verification TroubleShooting GuideFor In Vitro Diagnostic UseTroubleshooting21

Page 21 - Chapter 4: Shutdown

MAGPIX® User Quick Guide22For In Vitro Diagnostic Use

Page 22

On the WebFor more information, visit the Luminex FAQ page at http://www.luminexcorp.com/support/.You can access the Technical Support website using a

Page 23 - Troubleshooting

© Luminex Corporation, 2011. All rights reserved. No part of this publication may bereproduced, transmitted, transcribed, or translated into any langu

Page 24

MAGPIX® User Quick GuideiiFor In Vitro Diagnostic Use

Page 25

Table of ContentsChapter 1 Starting the MAGPIX System for the First Time ...1Starting xPONENT ...

Page 26

MAGPIX® User Quick GuideivFor In Vitro Diagnostic Use

Page 27 - Shooting Guide

Detailed instructions on how to install the MAGPIX System are provided in the PC InstallationInstructions (89-30000-00-263) and Installing MAGPIX (89-

Page 28

When you turn on the Luminex MAGPIX system, the following start up indicators occur:• Blue light indicator turns on.• Fans turn on.• Syringe pump init

Page 29 - Chapter 6: Technical Support

1. On the Home page, click Probe and Heater under Daily Activities. The Probe &Heater tab opens.2. Based on the type of plate you are using, place