MAGPIX® User QuickGuideFor In Vitro Diagnostic Use
CAUTION: Correct sample probe height is critical to successful sampleacquisition and calibration. Problems with the sample probe canlead to fluid leak
1. Open the Maintenance page, then the Cmds & Routines tab.2. Select Revive After Storage (Luminex) from the drop-down list. The Revive AfterStora
System InitializationSystem initialization prepares the system for data acquisition. On initial startup, the SystemInitialization should consist of:•
1. Load the CAL/VER CD (provided with the kit) on the computer.2. On the Home page, click System Initialization.3. Click Import Kit at the bottom righ
MAGPIX® User Quick Guide8For In Vitro Diagnostic Use
Defining the System Initialization RoutineNOTE: Luminex recommends daily performance verification and weeklycalibration of the MAGPIX system. You can
1. On the Home page, click System Initialization under Daily Activities. The Auto Mainttab opens.2. On the Auto Maint tab, the System Initialization o
Creating a KitTo create a kit:1. Open the Protocols page, then open the Protocols tab. Select the protocol that youwant to use for the kit, then click
1. Open the Protocols page, then open the Protocols tab. Click the Stds & Ctrls tab, thenclick Create New Std/Ctrl Lots.2. In the Select Protocol
5. If you want a description for the batch, type it in the Enter Optional Description box.• Click the existing protocol that you want to use. If the p
6. Click Generate. The report appears in the lower part of the Reports tab.7. Click Print to print the report, or Save to save the report as a PDF fil
Shutting Down the AnalyzerRun the daily shutdown routine to prevent clogs and crystallization of salt in the sampleprobe. Clogs and crystallization of
MAGPIX® User Quick Guide16For In Vitro Diagnostic Use
MAGPIX New System Initialization TroubleshootingThis section provides you with information to troubleshoot issues that may be encounteredduring the MA
1. Adjusting the Sample Probe Height2. Revive After Storage Routine (Luminex)3. System Initialization All system startup routines are run from the MAG
Problem Summary Resolution NotesIf Calibration continues to fail runthe following routine:Create routines in thexPONENT “Maintenance” Tab“Cmds & R
Problem Summary Resolution Notes(1x) Sanitize with 0.1N NaOH(reservoir C)(2x) Clean with 0.1N NaOH(reservoir C)(2x) Rinse (reservoir D)(1x) Alcohol Fl
FIGURE 7.MAGPIX Calibration and Performance Verification TroubleShooting GuideFor In Vitro Diagnostic UseTroubleshooting21
MAGPIX® User Quick Guide22For In Vitro Diagnostic Use
On the WebFor more information, visit the Luminex FAQ page at http://www.luminexcorp.com/support/.You can access the Technical Support website using a
© Luminex Corporation, 2011. All rights reserved. No part of this publication may bereproduced, transmitted, transcribed, or translated into any langu
MAGPIX® User Quick GuideiiFor In Vitro Diagnostic Use
Table of ContentsChapter 1 Starting the MAGPIX System for the First Time ...1Starting xPONENT ...
MAGPIX® User Quick GuideivFor In Vitro Diagnostic Use
Detailed instructions on how to install the MAGPIX System are provided in the PC InstallationInstructions (89-30000-00-263) and Installing MAGPIX (89-
When you turn on the Luminex MAGPIX system, the following start up indicators occur:• Blue light indicator turns on.• Fans turn on.• Syringe pump init
1. On the Home page, click Probe and Heater under Daily Activities. The Probe &Heater tab opens.2. Based on the type of plate you are using, place