Exporting a Routine
1. Open the Maintenance
page, then open the Cmds & Routines tab.
2. Click Export.
3. In the Save As dialog box, navigate to the folder where you want to store the routine file,
then click Save.
NOTE: The export option is only available for routines that have been saved.
Probe and Heater Tab
Use this tab to adjust the sample probe height and plate heater settings.
FIGURE 48.
Probe and Heater Tab
Auto Adjust Height - Automatically adjusts the probe height using the specified location(s) in
the main plate, reservoir, and strip wells.
NOTE: Luminex recommends that you calibrate the sample probe height
using well A1.
Performing System Maintenance
101