Page 1 - Guide 4.2

MAGPIX® User QuickGuide 4.2

Page 2

System InitializationSystem initialization prepares the system for data acquisition. On initial startup, the SystemInitialization should consist of:•

Page 3

1. Load the CAL/VER CD (provided with the kit) on the computer.2. On the Home page, click System Initialization.3. Click Import Kit at the bottom righ

Page 4 - Table of Contents

Defining the System Initialization RoutineNOTE: Luminex recommends daily performance verification and weeklycalibration of the MAGPIX system. You can

Page 5

3. Under Maintenance Options, select Fluidics prep, performance verification from thedrop-down list. FIGURE 2.Setting System Initialization Routine4.

Page 6 - Start-up Indicators

Creating a Quantitative Assay ProtocolThe protocol must contain multiple standards. The standards are assigned lot valueinformation for each test. A s

Page 7 - Initial Startup

15. In the Analysis Settings dialog box, select the analysis method from the Method list,and the weighting in the Weight Type list. Click Apply to All

Page 8

27. Click Next. The Plate Layout tab opens. • To add well commands, highlight the appropriate wells and mark them as unknown,standard, control, backgr

Page 9 - Revive After Storage Routine

default Post Batch routine can be changed in the Batch Optionstab of the Admin page.Creating a KitTo create a kit:1. Open the Protocols page, then ope

Page 10 - System Initialization

Create a New Batch from an Existing ProtocolThis option creates a new batch using a selected protocol from the Installed Protocols list.The list conta

Page 11

1. Open the Results page, then the Reports tab.2. In the Report drop-down list, select the category of report: batch, protocol, calibrationand verific

Page 12

© Luminex Corporation, 2011. All rights reserved. No part of this publication may bereproduced, transmitted, transcribed, or translated into any langu

Page 13

Shutting Down the AnalyzerRun the daily shutdown routine to prevent clogs and crystallization of salt in the sampleprobe. Clogs and crystallization of

Page 14 - Chapter 3: Running Assays

MAGPIX® User Quick Guide 4.216

Page 15

If you are not successful troubleshooting problems with MAGPIX, contact Luminex TechnicalSupport. See Technical Support. If necessary, attach a suppor

Page 16

FIGURE 3.MAGPIX Off-Plate Reagent Block1 Reservoirs2 Strip WellsMAGPIX New System Startup ErrorsProblem Resolution1. Check the sample probe height for

Page 17 - Creating a Lot

Problem Resolution(1x) Clean with 0.1 N NaOH (reservoir C)(3x) Rinse (reservoir D)Run Calibration routine. Verify that reagent is freshly vortexed. Us

Page 18 - Generating a Report

Problem ResolutionVER (strip well B)Verify that reagent is freshly vortexed. Use 6 drops of reagent.(1x) Rinse (reservoir D)Fluidics 1 (strip well C)V

Page 19

MAGPIX Calibration and Performance VerificationTroubleshooting GuideTroubleshooting21

Page 20 - Chapter 4: Shutdown

TABLE 1.Fittings to CheckFitting Location, Step, andFittingFluid CompartmentStep 3: Drive Fluid FittingFluid CompartmentStep 4: Waste ContainerFitting

Page 21 - User Quick Guide 4.2

You can access the Technical Support website using a user name and password athttps://oraweb.luminexcorp.com/OA_HTML/jtflogin.jsp.By Phone Inside the

Page 22 - Chapter 5: Troubleshooting

8. Verify the location where you want to store the file. To change the location, click Browse,then navigate to the new folder and click OK.9. Click Sa

Page 23 - FIGURE 3

MAGPIX® User Quick Guide 4.2ii

Page 24 - Troubleshooting

Table of ContentsChapter 1 Starting the MAGPIX System for the First Time ...1Starting xPONENT ...

Page 25 - Changing Sample Load Volume

MAGPIX® User Quick Guide 4.2iv

Page 26 - Troubleshooting Guide

Detailed instructions on how to install the MAGPIX System are provided in the PC InstallationInstructions (89-30000-00-263) and Installing MAGPIX (89-

Page 27 - Technical Support

• Blue light indicator turns on.• Fans turn on.• Syringe pump initializes.• The filter wheel initializes.• Software connects to the instrument.NOTE: I

Page 28 - Sending a Support.zip File

1. On the Home page, click Probe and Heater under Daily Activities. The Probe &Heater tab opens.2. Use well D6 (this is the center of a standard 9

Page 29

FIGURE 1.Sample Probe Height AdjustmentRevive After Storage RoutineNOTE: The Revive After Storage routine is necessary when the systemruns for the fir