Luminex FLEXMAP 3D User Manual Manuel d'utilisateur Page 30

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 86
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 29
CAUTION: Adhere to standard laboratory safety practices when handling
hazardous, toxic, or flammable reagents and chemicals. Contact
Luminex Technical Support when in doubt about compatibility of
cleaning and decontamination agents or materials.
Required Laboratory Reagents
10 - 20% household bleach
70% isopropanol or 70% ethanol
0.1N NaOH
Sporicidin
®
Disinfectant
Mild detergent
Deionized water
WARNING: Isopropanol and ethanol are flammable liquids. Keep them away
from heat, open flames, and sparks in a well-ventilated area.
Remove them from the instrument when they are not in use.
Luminex
®
xMAP
®
Technology Reagents
F3DCAL1 - Classification Calibrator (calibrates DD, CL1, CL2, CL3 - for all xMAP
®
microsphere applications other than MagPlex microspheres)
F3DeCAL1 - Classification Calibrator (calibrates DD, CL1, CL2, CL3 - for MagPlex
microsphere-based applications only)
F3DCAL2 - Reporter Calibrator (calibrates RP1 standard and enhanced range - for all
xMAP microspheres)
F3DCAL3 - EDR Calibrator (calibrates RP1 extended range - for all xMAP microspheres)
F3DVER1 - Classification Verifier (verifies DD, CL1, CL2, CL3 - for all xMAP microsphere-
based applications other than MagPlex microspheres)
F3DeVER1 - Classification Verifier (verifies DD, CL1, CL2, CL3 - for MagPlex microsphere-
based applications only)
F3DVER2 - Reporter Verifier (verifies RP1 standard and enhanced PMT and extended
range - for all xMAP microspheres)
Fluidics1 - xMAP Fluidics 1 (verifies fluidic integrity and well-to-well microsphere carryover)
Fluidics2 - xMAP Fluidics 2 (verifies fluidic integrity and well-to-well microsphere carryover)
Luminex xMAP Sheath Fluid
CAUTION: Protect xMAP reagents from light at all times to avoid
photobleaching of the microspheres.
CAUTION: Luminex reagents contain ProClin
®
as a preservative. This can
cause allergic reactions in some people.
Luminex
®
FLEXMAP 3D
®
Hardware User Manual
20
For research use only. Not for use
in diagnostic procedures.
Vue de la page 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 85 86

Commentaires sur ces manuels

Pas de commentaire