NOTE: The Luminex
®
FLEXMAP 3D
®
Probe Height Adjustment Tool is to be
used for 384 well mylar bottom and 384 well filter bottom plates only.
To calibrate the probe correctly, perform the following steps:
1.
Using data sheet for the plate you intend to calibrate the probe to, subtract the "well
depth" from the "plate height" as shown in the example below:
Plate Height: 14.2mm - Well Depth: 11.43mm = 2.77mm (2.8mm)
Example 1: 2.8mm or well C7 on the probe height adjustment tool for this example.
Appendix D: FLEXMAP 3D
®
Probe Height
Adjustment Tool