Detects and distinguishes surface reporter fluorescence emissions at 575 nm on the surface of 1 to
500 unique xMAP microspheres in a single sample
Soluble background fluorescence emission at 575 nm automatically subtracted from fluorescence
intensity values.
Microtiter Plates
The Luminex
®
FLEXMAP 3D
®
system processes a 96-well microtiter plate in less than 20
minutes and a 384-well microtiter plate within 1 hour and 15 minutes, with 2500 microspheres
per region per well, counting 100 microspheres in each region.
Microtiter plates with 96 and 384 wells must be compatible with the microtiter heater block
temperature (from 35 to 60°C or 95° to 131°F) when the heater block is in use.
All microtiter plates (96 or 384 wells) have standard width (85.6 mm) and length (127.9 mm).
Depth varies depending on the type of well. To be compatible with the
FLEXMAP 3D
®
instrument, the maximum allowable depth is 1”. Plates must have minimum 0.06” lip height,
standard distance from well center to well center (9 mm on 96-well plate), and standard
distance from A1 center to plate center in both length and width. To be compatible in size
with the microtiter heater block, the plate must fit into the heater block so that the top is flush
with the heater block.
TABLE .
Continued
For research use only. Not for use in diagnostic procedures.
Technical Overview
25