Luminex BioPlex XYP, LX200 User Manual (supplied by BioRad Manuel d'utilisateur Page 42

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 51
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 41
5.3 Technical Support
For technical assistance with the Bio-Plex 200 system, including all hardware and software,
contact your local Bio-Rad office or, in the US, call 1-800-424-6723. All accessories and spare
parts not listed in this document can be ordered similarly, or write to Bio-Rad Laboratories, Inc.,
2000 Alfred Nobel Drive, Hercules, CA 94547.
Section 6
Bio-Plex 200 System Specifications
General Technical Specifications
Environmental conditions
Operating temperature 15–30°C (59–86°F)
Operating humidity 20–80%, noncondensing
Operating altitude Designed to operate at 2,400 m (7,874 ft) above
mean sea level or below
Compensatory range ± 2°C
UL installation category UL Installation Category II, as defined in Annex J of
UL 3101-1
Pollution degree Pollution Degree 2, as defined in Section 3.7.3.2 of
UL 3101-1
Array Reader Specifications
Input voltage range 100–240 V, ~1.5 A, 47–63 Hz
Physical dimensions (W x D x H) 43 x 51 x 23 cm
Weight 23 kg (60 lb)
Lasers
Reporter laser 532 nm, 10 mW
Classification laser 635 nm, 8.5 mW, diode
Fluidics
Sheath flow rate 90 µl/sec
Cuvette 200 µm square flow channel
Sample injection rate 60 µl/min
Electronics
Reporter channel Photomultiplier tube, A/D resolution 14 bits
detection
Classification and Avalanche photodiodes with temperature
doublet discriminator compensation, A/D resolution 12 bits
channel detection
Communications interface RS232 and USB
Signal processing
Measurement resolution 15 bits effective
Processor modes Linear, with logarithmic or linear display option
Dynamic range 70 dB
36
www.bio-rad.com/bioplex/
Vue de la page 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 50 51

Commentaires sur ces manuels

Pas de commentaire