Batch Options - This section displays the following options:
Allow running or saving a batch with expired reagents and fluids - Allows batches
to run with expired reagents and fluids.
Allow running a batch if XY temperature is not in range - Allows batches to run if the
XY temperature is not in range.
Allow running a batch if the instrument is not calibrated or verified - Allows batches
to run if the instrument is not calibrated.
NOTE: Settings to enable running/saving with expired agents, running with
XY temperature out of range, and running a batch if the instrument is
not calibrated or verified are managed at the system level. If one or
more of these options is enabled, you cannot run batches while the
condition exists. However, if you have the group permission Batch
run override system, you can run a batch with the condition
present. Without Batch run override system you are unable to
override any or all of these conditions and therefore cannot run or
save batches.
Use Weighting as default option for quantitative analysis - Sets the weighting for
quantitative analysis to default to ON for new protocols.
Warn when saving/running a New Batch with New Protocol without saving the
Protocol and/or Standard and Control information -Opens a dialog box warning
when saving or running a new batch with a new protocol when you haven’t saved the
protocol and/or standard and control information.
Prompt for reacquiring of the last errored wells for an aborted partial batch -
Allows users to choose to reacquire aborted wells when resuming a stopped/partial
batch or to start with the next unacquired well.
Batch Settings - This section displays the following options:
Calibration expiration days - Type the desired number of days a system calibration
remains active before expiring. The default number of days is seven.
Analysis display decimal places - Type the number of digits the system displays. The
default number of digits is three.
Minimum bead count for obtaining results - Type any whole number from 0 to
1000000. This is the minimum bead count that must be observed by the instrument for a
particular analyte before this data is used in statistical calculations and is displayed on
the graphs and results table. When set to a number greater than 0, the analyzer does
not display data for bead sets not generating events equal to or less than this value.
Type 0 to display all events. The default setting is 1.
Allow batches to be run or saved without lot number, expiration or manufacturer -
Allows saving lots without normal required information.
Default Pre-Batch Routine - System maintenance routines like wash, flush, or clean
can be programmed at different points in a batch to prevent clog formation and maintain
peak system performance. Pre-batch routines, specifically, are recommended for
challenging sample matrices like viscous material or material containing particulates in
suspension because these matrices can clog the instrument. Select this option to specify
a particular routine to run before running a batch.
NOTE: A Rinse command runs after all pre-batch routines.
Default Post-Batch Routine - A Luminex Post-Batch routine automatically runs at the
end of each batch to clean the sample chamber and maintain peak system performance.