Name box and box for description - Type a name and a description in the appropriate
boxes.
Acquisition Settings - Use this section to set the following:
Volume - This is the volume the instrument will aspirate into the system for analysis.
Type the desired sample volume in microliters. Use values from 20 to 200 µl. To avoid
air intake, add at least 25 µl to the sample well in addition to the sample size. The
default value is 50 µl.
XY heater - Select Enabled to enable the XY heater. In the box, type the desired value
in degrees Celsius. The temperature range is 35 to 60°C in 0.5 increments.
CAUTION: Acquiring data before the heater has reached the proper
temperature can compromise test results.
Plate Name - The name assigned to the plate during sample probe height adjustment.
You can select a different plate from the list.
Sample wash - Select this option for assays without a final wsh step prior to reading the
plate on the instrument. This automatically washes each sample within the instrument.
Final washes are required for proper analysis.