5. Remove the filter with one hand and hold the tubing with the other hand.
CAUTION: Do not allow the tubing to fall inside the instrument.
6. Connect a new filter to the tubing and position the filter inside the panel.
7. Reattach the panel door to the unit and close the access door.
8. Plug in the power cord and turn on the FLEXMAP 3D instrument.
FIGURE 17.
HEPA Air Filter
Cleaning the Ventilation Filters
Each filter has at least one imprinted arrow on its metal frame. This arrow indicates the
direction of the air flow. Filters should be installed so that the arrows are at the correct
locations and point in the correct direction.
To clean the Luminex
®
FLEXMAP 3D
®
ventilation filters:
1. Turn off the FLEXMAP 3D instrument and unplug the power cord.
2. Remove the XY cover by gently grasping its ventilation holes and pulling it straight out.
3. Slide clamps off the silver filter inside the cover and remove the filter.
NOTE: For filter location, reference the "Ventilation Filter on XY Platform"
image below.
4. Open the front doors. Slide clamps off the left and right silver door filters and remove the
filters.
NOTE: For clamp location, reference the "Ventilation Filters inside Doors"
image below.
5. Clean the filters with a vacuum or with distilled water.
FLEXMAP 3D
®
Hardware User Manual
30