FIGURE 22.
Normal Bead Detail
TABLE 9.
Bead Detail Irregularities
Symptom Description Possible Problem Solution
xMAP and calibrator microspheres form a group in the wrong location.
Calibrator
microspheres
group together at
the top of the
region.
The calibrator
microspheres are
photobleached.
Replace the calibrator microspheres
with unphotobleached calibrator
microspheres. To avoid photobleaching,
protect microspheres from light.
xMAP
microspheres
group together at
the lower right of
the region.
The xMAP
microspheres are
photobleached.
Replace the xMAP microspheres with
unphotobleached xMAP microspheres.
To avoid photobleaching, protect
microspheres from light.
xMAP microspheres form an unusual shape.
xMAP
microspheres
form a long
diagonal line.
The xMAP
microspheres have
agglutinated.
Contact Technical Support.
xMAP
microspheres
form a group with
extending tails.
The solvent is
incompatible.
The Luminex Technical Support website
(http://www.luminexcorp.com/Support)
lists incompatible solvents. If the solvent
you are using is listed, switch solvents.
xMAP microspheres fail to form a group.
xMAP
microspheres do
not form a
cohestive group.
The sheath fluid is
incompatible.
Use only Luminex xMAP sheath fluid in
the FLEXMAP 3D instrument. Other
fluids can damage the instrument and
void your warranty.
Troubleshooting
45