Symptom Possible cause Solution
There is air in the instrument. The
waste line was moved during
instrument operation, resulting in
an unstable flow rate.
Perform an automatic sample
probe height adjustment. Use the
software to run a prime command
three times, an alcohol flush
command twice, then a wash
command three time with distilled
water.
The waste line was moved during
instrument operation, resulting in
an unstable flow rate.
Stabilize the waste line during
instrument operation. See the
section titled “Maintaining Fluids”.
There is a problem internal to the
instrument.
Review the log of calibration
reports. Check for dramatic
changes in temperature, sheath
pressure, or voltage. If any of
these are present, contact
Technical Support.
No events are collected
during calibration.
There is a problem with fluid
levels.
Check the sheath and waste fluid
levels. Verify that tubing for both
bottles is tightly connected to the
instrument. Check that the waste
fluid container cap is vented.
The sample probe is clogged. See the section titled "Clogs".
The Cheminert fitting is loose. Tighten the Cheminert fitting.
Verification Problems
TABLE 7.
Verification Problems
Symptom Possible cause Solution
The instrument fails
verification.
The verification microspheres are
not fully suspended.
Vortex the verification vials to
resuspend the microspheres.
The wrong lot number or target
values are entered in the
appropriate page of the software
UI.
Check the lot number and target
values to make sure they are
correct.
The instrument verification
microspheres are in the wrong
well.
Verify that verification
microspheres are in the correct
well.
There are not enough verification
microspheres in the well.
Add at least five drops of
verification microspheres to the
well. For accurate drop volume,
hold the vial upside down at a 90
degree angle to the microtiter plate
while dispensing them.
FLEXMAP 3D
®
Hardware User Manual
42