Luminex xPONENT for MAGPIX 4.2 Manuel d'utilisateur Page 101

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 164
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 100
Analysis Type - Use this list to choose from the following analysis types:
None - No analysis. Select if you have your own data post-processing program. You
cannot apply standards or controls when you select None. You cannot analyze
acquisitions with this setting.
Qualitative - Qualitative analysis determines results as either positive or negative,
reactive or non-reactive. The software is flexible in defining custom results ranges, such
as negative, low positive, or high positive. Determinations are based on a single
standard. For qualitative analysis the Luminex software uses a specific algorithm, shown
below:
Where:
FI = Fluorescent Intensity
Ki - a "Quali" value determines a cutoff or threshold. This, in conjunction with ranges
using the Lum Qual formula or an edited range specific for your assay, helps to
determine qualitative results for unknown samples.
Two predefined formulas using the algorithm are included in the system. You can use
them as-is or edit their range values to meet your requirements.
Quantitative - Determines the sample concentrations from standard curves using
regression methods. Cublc Spline, Linear, Logistic 4P, and Logistic 5P. Type the desired
values for standards and controls in the Number of Standards and Number of
Controls boxes. Select either Fit of All Standards or Mean of Replicates for the
calculation of the curve fit.
NOTE: Luminex recommends Fit of All Standards as the most accurate
calculation of the curve fit.
Based on a range of quantitative, numerical results, a threshold range can be applied to
a quantitative analysis, for example, high, low, saturated, and expected.
Allele Call - Sets analysis for an allele call. Analytes must be put in groups of 2, 3, or 4.
Min MFI Enabled - Select this box to enable a minimum MFI for the Allele Call analysis.
Type a value in this box to set the minimum MFI for analysis.
Analyze results while acquiring samples - The software allows real-time viewing of the
results as the instrument analyzes samples. This feature is not available if you select None
as your analysis type.
Number of Standards - Click to type the number of standards for the protocol. Applies
only to qualitative and quantitative analysis.
Use External Analysis Program - Select this check box to use a third-party program to
analyze the data. The Analysis Program list becomes active when this is selected.
Applies only to qualitative and quantitative analysis. Use this list to select which program to
use for data analysis
Number of Controls - Click to type the number of controls for the protocol. Applies only to
qualitative and quantitative analyses.
xPONENT for MAGPIX 4.2 Software User Manual
88
Vue de la page 100
1 2 ... 96 97 98 99 100 101 102 103 104 105 106 ... 163 164

Commentaires sur ces manuels

Pas de commentaire